Detection and management of subclinical atrial fibrillation in implantable and wearable devices

Review Article
Issue
2021/06
DOI:
https://doi.org/10.4414/cvm.2021.02180
Cardiovasc Med. 2021;24:w02180

Published on 01.11.2021

Introduction

Atrial fibrillation (AF) is the most common cardiac arrhythmia and is associated with increased mortality, heart failure and stroke [1–3]. It is predicted that AF prevalence in Europe will more than double by the year 2060 [4]. Cardiac embolism due to atrial fibrillation accounts for one third of all ischaemic strokes [5, 6] and about 18% of all ischaemic strokes due to AF are related to subclinical AF (SCAF) [7]. Therefore, early detection of SCAF may be crucial. SCAF, sometimes also referred to as "silent" or "asymptomatic" AF, describes clinically unapparent AF, mostly of short duration, which is detected with an implanted device such as a pacemaker or internal cardioverter-defibrillator (ICD), or a wearable device such as a smartwatch. Atrial high rate episodes (AHREs) in pacemakers and ICDs need to be confirmed by a physician who visually reviews the intracardiac electrograms to be labelled as SCAF. The same applies to the ECG-recorded rhythm in wearables such as smartwatches [8]. Although SCAF is associated with stroke, it is important to note that the increase in risk appears to be lower when compared with symptomatic AF [9, 10]. Stroke prevention by the use of oral anticoagulation in patients with symptomatic AF is well established, but there are no data from randomised trials addressing the question whether patients with SCAF should receive oral anticoagulation, and very few patients with SCAF have been included in AF anticoagulation trials. Recently, numerous wearable devices to detect clinical and subclinical AF have become available. As a result, technological development may lead to a shift from physician-initiated to patient-initiated screening for AF.

Screening for AF?

Even if very recently, results from the STROKESTOP trial have reported a small clinical benefit in the elderly population when they participated in a screening programme for AF (results presented by Emma Svennberg during EHRA congress on 23 April 2021), the role of systematic screening in the general population remains uncertain and is therefore not recommended by current guidelines. Nevertheless, opportunistic screening by pulse palpation followed by a 12-lead ECG in the case of an irregular pulse has been recommended for patients older than 65 years in European Society of Cardiology (ESC) guidelines since 2012 [11], as well as in the 2020 version of the AF guidelines [8]. In patients with a history of an ischaemic stroke or systemic embolism, secondary screening for AF is challenging because paroxysmal AF episodes often remain undetected. AF detection is related to total AF burden and obviously improves with increasing intensity of monitoring [12]. As an example, wearing a cardiac event monitor belt for 30 days results in higher yield of AF detection than standard 24-hour Holter monitoring as shown in the EMBRACE trial (30-Day Cardiac Event Monitor Belt for Recording Atrial Fibrillation After a Cerebral Ischemic Event) [13]. In patients with cryptogenic stroke, intensified monitoring with an implantable loop recorder leads to an increase in AF detection when compared with standard care: 30% versus 3% after 36 months in the CRYSTAL-AF (Cryptogenic Stroke and Underlying Atrial Fibrillation) trial [14]. Implantable loop recorders are safe and effective tools for detecting AF episodes and the number of implantations of such devices is increasing as a result of technological improvements including miniaturisation [15]. However, when elderly asymptomatic patients are intensively screened for AF one may anyway find a high AF incidence. This was examined in the ASSERT II study (Prevalence of Sub-Clinical Atrial Fibrillation Using an Implantable Cardiac Monitor), where an annual incidence of SCAF lasting >5 minutes was found in 34.4% of elderly asymptomatic patients without pacemakers who were implanted with a loop recorder [16].

AF detection using cardiac implantable electronic devices

There is no better heart rhythm monitor than a pacemaker or ICD with an atrial lead, since it provides continuous monitoring. Therefore, the detection of SCAF in patients with cardiac implanted electronic devices (CIEDs) is frequent. Prevalence ranges between 25% and 50% during the first 1 to 3 years [10, 17–20]. Several trials have investigated the relationship of SCAF detected in CIEDs and the association with stroke. It is important to note that in patients with continuous monitoring by CIED there is often no strict temporal relationship between AF episodes and occurrence of stroke, suggesting that AF may be a marker rather than a causal factor of stroke risk [17, 21, 22]. In the ASSERT trial (Asymptomatic Atrial Fibrillation and Stroke Evaluation in Pacemaker Patients and the Atrial Fibrillation Reduction Atrial Pacing Trial) [10] 10.1% of device patients experienced at least one episode of SCAF >6 minutes duration in the first 3 months. Stroke rate was elevated 2.5 times compared with patients without subclinical AF but this was lower than stroke rate in patients with clinical, symptomatic AF. In the MOST (Mode Selection Trial) study atrial high rate episodes >5 minutes were associated with increased stroke risk [9]. The investigators of the TRENDS study (A Prospective Study of the Clinical Significance of Atrial Arrhythmias Detected by Implanted Device Diagnostics) found that an increased AF burden (>5.5 hours during 30 days) was associated with stroke [17]. In summary, the risk of stroke in patients without anticoagulation increases with SCAF duration and burden, and the strongest evidence for a benefit of anticoagulation was noted when episodes >24 hours duration were detected [23]. However, a clear threshold when to start with oral anticoagulation is not well defined.

Management of subclinical AF

In contrast to symptomatic AF, evidence on appropriate management of SCAF is limited [12]. In clinical trials, SCAF was associated with increased mortality and stroke when lasting longer than 5 minutes [9, 24] or 6 minutes [10], and in particular when lasting >24 hours [25, 26]. The current ESC AF guidelines state that available evidence is insufficient to justify routine use of oral anticoagulation in patients with SCAF, and modifiable risk factors should be identified and managed in each patient [8]. Patients with SCAF observed during device follow-up should be regularly monitored for progression to clinical AF and changes in individual thrombo-embolic risk. In patients with longer SCAF episodes, especially >24 hours, and a high CHA2DS2VASC-score (≥2 in men or ≥3 in women) it is reasonable to consider the use of oral anticoagulation according to the guidelines [8]. A proposed management approach shown in figure 1.
Figure 1: 
Proposed management of AHRE/SCAF. AF = atrial fibrillation; AHRE = atrial high-rate episode; CKD = chronic kidney disease; CHA2DS2-VASc = Congestive heart failure, Hypertension, Age ≥75 years, Diabetes mellitus, Stroke, Vascular disease, Age 65–74 years, Sex category (female); f = female; LA = left atrium; LoE = level of evidence; m = male; OAC = oral anticoagulant; SCAF = subclinical atrial fibrillation. 
(a Highly selected patients (e.g., with previous stroke and/or age ≥75 years, or ≥3 CHA2DS2-VASc risk factors, and additional non-CHA2DS2-VASc stroke factors such as CKD, elevated blood biomarkers, spontaneous echo contrast in dilated LA, etc); selected patients (e.g., with previous stroke and/or age ≥75 years, or ≥3 CHA2DS2-VASc risk factors , etc).

AF detection in wearable devices

Wearables are smart electronic devices worn close to the skin such as smartwatches, chest straps and ECG patches with connectivity to a smartphone. In comparison to monitoring with Holter ECGs, which usually cover a monitoring period between 24 hours up to 7 days, wearables provide non-invasive cardiac rhythm monitoring without limitation of time. SCAF may occur very infrequently. As an example, mean time to detection of SCAF (by CIED) in the ASSERT trial was 36 days [10] and in ASSERT II (by implantable loop recorder) was 5.1 ± 5.5 months. Hence, this is a possible gap where noninvasive monitoring using wearables might be beneficial. Although wearables such as chest strap monitors and ECG patches provide continuous monitoring, they may be less comfortable to wear and recently smartwatches have become more and more popular in the general population. Importantly, smartwatches do not provide continuous rhythm monitoring because of battery limitations. Instead they spot check heart rate and heart rhythm through photoplethysmography (PPG). The PPG technique measures changes in the microvascular blood volume, which converts into pulse wave and tachogram recording. This is achieved by using an emitter that sends a continuous pulse through the skin and a photodetector that measures the variable intensity of reflected photons from the tissue [27]. PPG records are affected by motion artefacts, pressure disturbances and skin pigmentation, but this can be improved using signal processing techniques [27]. Of note, current smartwatch algorithms are not able to detect AF below a heart rate of 50 bpm or above a heart rate of 120 bpm [28], which is a limitation. Smartwatch algorithms notify the user to obtain a single-lead ECG, which resembles lead I, for a duration of 30 seconds as soon as an irregular heart rhythm is detected. To obtain a single-lead ECG, smartwatch users can place a contralateral finger on the crown (negative electrode on the side of the watch), with the back of the watch serving as the positive electrode [29]. Some smartwatches are already officially labelled as medical devices. Recently, the Withings Move© smartwatch has received CE marking, whereas the Apple Watch©, Samsung Galaxy Watch© and Fitbit Sense© have received CE marking as well as FDA approval for detection of sinus rhythm or AF.
Two large studies with smartwatches were conducted in the general population using the photoplethysmography (PPG) method – the Apple Heart Study [30] and the Huawei Heart Study [31]. In the Huawei Heart Study, the mean age of the participants was 34.7 ± 11.5 years, with only 6% being older than 54 years. In the Apple Heart Study, 15.9% of participants were older than 54 years. In the Apple Heart Study 0.5% of participants received a notification that an irregular rhythm was obtained by the PPG signal. Finally, AF was confirmed in 34% of those patients who did further testing with an ECG patch. In the Huawei Heart Study, 0.2% had an irregular rhythm and 87% of those that underwent follow-up with physical examination, ECG and 24-hour Holter were finally diagnosed with AF. These data emphasise that we have to be cautious using this technique, since the population using smartwatches is generally younger than the group at risk but SCAF episodes with unclear clinical implications might still be revealed. The use of wearables may lead to increasing number of requests to physician from patients presenting with a rhythm strip. Using the smartwatch technology in a population other than in patients at risk may probably increase the risk of false-positive findings in a presumably healthy population [28]. In summary, studies have shown the feasibility to detect AF in smartwatches, however it is still necessary that a physician approves the ECG recorded rhythm. Nevertheless, smart wearables are not a continuous heart rhythm monitor, rather a “spot check” of arrhythmias and currently do not replace the continuous monitors that are prescribed by a healthcare provider [32].

Outlook

Cardiologists play a major role in interpreting electrograms and rhythm strips and in providing guidance towards optimal therapy including anticoagulation. To date, there is still a major knowledge gap in what to do in patients presenting with SCAF recorded by CIED with a duration between 5 or 6 minutes and 24 hours. Even if SCAF duration >5–6 minutes in combination with an elevated CHA2DS2VASC score is associated with an increased stroke risk, the risk-benefit ratio of starting anticoagulation in SCAF <24 hours remains unclear [12]. Randomised trials addressing this grey area are ongoing. The NOAH-AFNET 6 trial (Non-vitamin K Antagonist Oral Anticoagulants in Patients With Atrial High Rate Episodes) will investigate whether patients with device-detected atrial high rate episodes/ SCAF >6 minutes will benefit from anticoagulation with edoxaban [33]. The ARTESIA study (Apixaban for the Reduction of Thrombo-Embolism in Patients With Device-Detected Sub-Clinical Atrial Fibrillation) will determine whether oral anticoagulation therapy with apixaban compared with aspirin reduces the risk of stroke or systemic embolism in patients with SCAF lasting between 6 minutes and 24 hours [34]. While we are waiting for these trial results, the authors of this review do not anticoagulate patients with device-detected SCAF <6 minutes duration according to available data, whereas in SCAF >24 minutes anticoagulation is started when indicated according to the CHA2DS2VASC score. In the grey zone in between, a clinical scenario that often occurs, the authors tend towards anticoagulation when a high CHA2DS2VASC score (≥2 in men, ≥3 in women) is indicative of increased stroke risk and especially when there is no history of bleeding. We recommend a visual review of the electrograms of atrial high rate episodes on a regular base to determine if SCAF is present, assess AF burden, look for a possible progression to clinical AF in which anticoagulation is better established, and to identify and treat comorbidities in AF.
The increasing use of smartwatches comes with opportunities and challenges. Cardiologists will probably encounter an increasing number of patients presenting with suspicion of AF. A visually confirmed episode of clinical AF detected by a smartwatch is already an indication for starting oral anticoagulation when thromboembolic risk is elevated according to the CHA2DS2VASC score, but there is a lack of data on whether anticoagulation in patients with SCAF diagnosed by a smartwatch is beneficial. The HEARTLINE trial (A Heart Health Study Using Digital Technology to Investigate if Early AF Diagnosis Reduces the Risk of Thromboembolic Events Like Stroke IN the Rea-world Environment) is investigating whether detecting symptomatic or asymptomatic AF with a smartwatch will improve clinical outcomes in people aged 65 and older when embedded in an app-based program including tools to promote adherence to oral anticoagulation. Meanwhile, we recommend dealing with SCAF recorded by a smartwatch in the same way that we deal with SCAF recorded by a CIED.
Cardiologists and other healthcare providers will also play a major role in guiding patients who desire to monitor their heart rhythm with wearables. Smartwatches could potentially fill the diagnostic gap between Holter monitoring and implantation of an implantable loop recorder in patients at increased risk of stroke. However, we opt against AF screening with smartwatches in the general population, and especially in younger, healthy patients given the low prevalence of AF in this population, the risk of false positive results and the possibility of gaining the diagnosis of a cardiac disorder with unclear treatment implications.

Key points

  • Implantable loop recorders (ILR) and cardiac implantable electronic devices (CIEDs) are considered to have the highest sensitivity to detect atrial fibrillation (AF).
  • Wearables such as smartwatches are increasingly used and will presumably increase the number of patients diagnosed with subclinical AF.
  • AF episodes detected by implanted or wearable devices need to be confirmed by visually reviewing the intracardiac electrogram or ECG-recorded rhythm.
  • Current data does not support anticoagulation in subclinical device-detected AF of < 5–6 minutes duration but does support anticoagulation when the duration is >24 hours
  • Ongoing studies such as the ARTESIA (Apixaban for the Reduction of Thrombo-Embolism in Patients With Device-Detected Sub-Clinical Atrial Fibrillation) trial examine the benefit of anticoagulation in the grey zone between 5–6 minutes and 24 hours
No financial support and no other potential conflict of interest relevant to this article was reported.
Prof. Laurent M. Haegeli, MD

Division of Cardiology

Medical University Department

Kantonsspital Aarau

Tellstrasse 25

CH-5001 Aarau


laurent.haegeli[at]ksa.ch
1. . Atrial fibrillation: a major contributor to stroke in the elderly. The Framingham Study. Arch Intern Med. 1987 Sep;147(9):1561–4. http://dx.doi.org/10.1001/archinte.1987.00370090041008 PubMed 0003-9926
2. . A population-based study of the long-term risks associated with atrial fibrillation: 20-year follow-up of the Renfrew/Paisley study. Am J Med. 2002 Oct;113(5):359–64. http://dx.doi.org/10.1016/S0002-9343(02)01236-6 PubMed 0002-9343
3. . Increased atrial fibrillation mortality: united States, 1980-1998. Am J Epidemiol. 2002 May;155(9):819–26. http://dx.doi.org/10.1093/aje/155.9.819 PubMed 0002-9262
4.  Projections on the number of individuals with atrial fibrillation in the European Union, from 2000 to 2060. Eur Heart J. 2013 Sep;34(35):2746–51. http://dx.doi.org/10.1093/eurheartj/eht280 PubMed 1522-9645
5.  Adelaide stroke incidence study: declining stroke rates but many preventable cardioembolic strokes. Stroke. 2013 May;44(5):1226–31. http://dx.doi.org/10.1161/STROKEAHA.113.675140 PubMed 1524-4628
6. . High prevalence of atrial fibrillation among patients with ischemic stroke. Stroke. 2014 Sep;45(9):2599–605. http://dx.doi.org/10.1161/STROKEAHA.114.006070 PubMed 1524-4628
7.  Subclinical device-detected atrial fibrillation and stroke risk: a systematic review and meta-analysis. Eur Heart J. 2018 Apr;39(16):1407–15. http://dx.doi.org/10.1093/eurheartj/ehx731 PubMed 1522-9645
8. . 2020 ESC Guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS): The Task Force for the diagnosis and management of atrial fibrillation of the European Society of Cardiology (ESC) Developed with the special contribution of the European Heart Rhythm Association (EHRA) of the ESC. Eur Heart J. 2021 Feb;42(5):373–498. http://dx.doi.org/10.1093/eurheartj/ehaa612 PubMed 1522-9645
9. . Atrial high rate episodes detected by pacemaker diagnostics predict death and stroke: report of the Atrial Diagnostics Ancillary Study of the MOde Selection Trial (MOST). Circulation. 2003 Apr;107(12):1614–9. http://dx.doi.org/10.1161/01.CIR.0000057981.70380.45 PubMed 1524-4539
10. . Subclinical atrial fibrillation and the risk of stroke. N Engl J Med. 2012 Jan;366(2):120–9. http://dx.doi.org/10.1056/NEJMoa1105575 PubMed 1533-4406
11. . 2012 focused update of the ESC Guidelines for the management of atrial fibrillation: an update of the 2010 ESC Guidelines for the management of atrial fibrillation—developed with the special contribution of the European Heart Rhythm Association. Europace. 2012 Oct;14(10):1385–413. http://dx.doi.org/10.1093/europace/eus305 PubMed 1532-2092
12. . Screening for atrial fibrillation: a European Heart Rhythm Association (EHRA) consensus document endorsed by the Heart Rhythm Society (HRS), Asia Pacific Heart Rhythm Society (APHRS), and Sociedad Latinoamericana de Estimulación Cardíaca y Electrofisiología (SOLAECE). Europace. 2017 Oct;19(10):1589–623. http://dx.doi.org/10.1093/europace/eux177 PubMed 1532-2092
13. . Atrial fibrillation in patients with cryptogenic stroke. N Engl J Med. 2014 Jun;370(26):2467–77. http://dx.doi.org/10.1056/NEJMoa1311376 PubMed 1533-4406
14. . Cryptogenic stroke and underlying atrial fibrillation. N Engl J Med. 2014 Jun;370(26):2478–86. http://dx.doi.org/10.1056/NEJMoa1313600 PubMed 1533-4406
15. . The Role of Implantable Cardiac Monitors in Atrial Fibrillation Management. J Atr Fibrillation. 2017 Aug;10(2):1590. http://dx.doi.org/10.4022/jafib.1590 PubMed 1941-6911
16. . Subclinical Atrial Fibrillation in Older Patients. Circulation. 2017 Oct;136(14):1276–83. http://dx.doi.org/10.1161/CIRCULATIONAHA.117.028845 PubMed 1524-4539
17.  The relationship between daily atrial tachyarrhythmia burden from implantable device diagnostics and stroke risk: the TRENDS study. Circ Arrhythm Electrophysiol. 2009 Oct;2(5):474–80. http://dx.doi.org/10.1161/CIRCEP.109.849638 PubMed 1941-3084
18.  Diagnosis of paroxysmal atrial fibrillation in patients with implanted pacemakers: relationship to symptoms and other variables. Pacing Clin Electrophysiol. 2009 Jan;32(1):91–8. http://dx.doi.org/10.1111/j.1540-8159.2009.02181.x PubMed 1540-8159
19. . Clinical Implications of Brief Device-Detected Atrial Tachyarrhythmias in a Cardiac Rhythm Management Device Population: Results from the Registry of Atrial Tachycardia and Atrial Fibrillation Episodes. Circulation. 2016 Oct;134(16):1130–40. http://dx.doi.org/10.1161/CIRCULATIONAHA.115.020252 PubMed 1524-4539
20.  Presence and duration of atrial fibrillation detected by continuous monitoring: crucial implications for the risk of thromboembolic events. J Cardiovasc Electrophysiol. 2009 Mar;20(3):241–8. http://dx.doi.org/10.1111/j.1540-8167.2008.01320.x PubMed 1540-8167
21. . Atrial fibrillation burden and atrial fibrillation type: clinical significance and impact on the risk of stroke and decision making for long-term anticoagulation. Vascul Pharmacol. 2016 Aug;83:26–35. http://dx.doi.org/10.1016/j.vph.2016.03.006 PubMed 1879-3649
22.
23.  Practice Variation in Anticoagulation Prescription and Outcomes After Device-Detected Atrial Fibrillation. Circulation. 2019 May;139(22):2502–12. http://dx.doi.org/10.1161/CIRCULATIONAHA.118.038988 PubMed 1524-4539
24.  Newly detected atrial high rate episodes predict long-term mortality outcomes in patients with permanent pacemakers. Heart Rhythm. 2014 Dec;11(12):2214–21. http://dx.doi.org/10.1016/j.hrthm.2014.08.019 PubMed 1556-3871
25. . Early detection of atrial high rate episodes predicts atrial fibrillation and thromboembolic events in patients with cardiac resynchronization therapy. Heart Rhythm. 2015 Dec;12(12):2368–75. http://dx.doi.org/10.1016/j.hrthm.2015.07.007 PubMed 1556-3871
26.  Duration of device-detected subclinical atrial fibrillation and occurrence of stroke in ASSERT. Eur Heart J. 2017 May;38(17):1339–44. http://dx.doi.org/10.1093/eurheartj/ehx042 PubMed 1522-9645
27. . Sensors and Functionalities of Non-Invasive Wrist-Wearable Devices: A Review. Sensors (Basel). 2018 May;18(6):18. http://dx.doi.org/10.3390/s18061714 PubMed 1424-8220
28. . Atrial fibrillation monitoring with wrist-worn photoplethysmography-based wearables: state-of-the-art review. Cardiovascular Digital Health Journal. 2020;1(1):45–51. http://dx.doi.org/10.1016/j.cvdhj.2020.03.001 2666-6936
29.  Smart wearable devices in cardiovascular care: where we are and how to move forward. Nat Rev Cardiol. 2021 Aug;18(8):581–99. http://dx.doi.org/10.1038/s41569-021-00522-7 PubMed 1759-5010
30. . Large-Scale Assessment of a Smartwatch to Identify Atrial Fibrillation. N Engl J Med. 2019 Nov;381(20):1909–17. http://dx.doi.org/10.1056/NEJMoa1901183 PubMed 1533-4406
31. . Mobile Photoplethysmographic Technology to Detect Atrial Fibrillation. J Am Coll Cardiol. 2019 Nov;74(19):2365–75. http://dx.doi.org/10.1016/j.jacc.2019.08.019 PubMed 1558-3597
32.  Smart Wearables for Cardiac Monitoring-Real-World Use beyond Atrial Fibrillation. Sensors (Basel). 2021 Apr;21(7):21. http://dx.doi.org/10.3390/s21072539 PubMed 1424-8220
33.  Probing oral anticoagulation in patients with atrial high rate episodes: rationale and design of the Non-vitamin K antagonist Oral anticoagulants in patients with Atrial High rate episodes (NOAH-AFNET 6) trial. Am Heart J. 2017 Aug;190:12–8. http://dx.doi.org/10.1016/j.ahj.2017.04.015 PubMed 1097-6744
34.  Rationale and design of the Apixaban for the Reduction of Thrombo-Embolism in Patients With Device-Detected Sub-Clinical Atrial Fibrillation (ARTESiA) trial. Am Heart J. 2017 Jul;189:137–45. http://dx.doi.org/10.1016/j.ahj.2017.04.008 PubMed 1097-6744